Autoclave use precautions

Autoclave use precautions

1, routine maintenance

1.1 The appearance of the sterilizer and the sterilization chamber should be kept clean and dry.

1.2 The probe and water level gauge should be cleaned regularly.

1.3 The door frame and rubber ring are not damaged, and the inlet port should not be blocked. It is best to apply talcum powder to the strip after daily use. To extend the life of the strip.

The interlocking device of the 1.4 door should be flexible and reliable, and can be opened freely.

1.5 The steam trap is cleaned once a month to facilitate the cooling of air and maintain the temperature.

2. After one year of use of the maintenance cycle, the qualified inspection department should be inspected annually for a comprehensive system inspection. Including: simplified, door, piping systems, electrical systems, etc. Safety valves, thermometers and pressure gauges should be regularly checked to ensure the safety and normal use of the equipment.

3. If the parking is long-term parking, the equipment should be placed in a ventilated and dry place, and should not be exposed to rain. If necessary, there should be coverings. The water in the steam generator should be drained and the door should be open and the chamber should be kept clean and dry.

4, daily use precautions

4.1 The sterilizer shall be operated by trained personnel, and the entire sterilization process shall be supervised by a person.

4.2 Can not completely rely on automatic water level protection, should always pay attention to the water level, so as not to burn the electric heating tube.

4.3 When adding water manually. The power should be cut off first, the vent valve 3 should be opened to relieve pressure, and then the inlet valve should be opened to add water. Never open the inlet valve when there is pressure on the interlayer. The vent valve should be open when adding water.

4.4 When there is pressure in the sterilization chamber, the interlock handle cannot be lifted, and the door cannot be forcibly opened.

4.5 Slowly release steam after sterilizing the liquid. The door can only be opened when the liquid temperature drops below 700C. Open the door immediately after sterilization is prohibited.

4.6 If there is a power failure or other reasons below the sterilization temperature during the sterilization process, the timing should be restarted when the temperature reaches the sterilization temperature again.

4.7 Pull the safety valve handle several times a week when there is pressure on the interlayer to ensure that the safety valve is working properly.

4.8 After the daily shift, remove the sterilizer's storage board, dry it with a cloth, and keep the sterilization room dry.

4.9 After the daily shift, the water pressure in the steam generator is discharged to reduce the scale.

4.10 The equipment should not be exposed during operation. When it is necessary to open, it is necessary to build a protective frame to prevent rain.

4.11 Continuous use time should be less than 8 hours / day.

Home and Office Table series are big series and multipurpose that it includes Side Table , Laptop Table , Computer Table and so on.Those can be used for Office and Home.Material also can be wood, MDF, steel frame and so on different on different customer`s requirement.

HEALTH AND WELLNESS: Standing Desk encourages greater movement throughout your day while helping optimize your cognition, improve focus, enhance collaboration, reduced back and neck pain, increase productivity and boost calorie burn.

Home Office Computer Table

Office Desk, Executive Desk, Standing Desk,Home Office Computer Table

Foshan Hollin Furniture Co.,Ltd , https://www.china-hollin.com