Instructions for use of Rubella enzyme immunoassay kit

This kit is for research use only.
Detection range: 96T
1 IU / ml -40 IU / ml

purpose of usage:
This kit is used to determine the content of rubella in monkey serum, plasma and related liquid samples.
Experimental principle This kit uses the double antibody sandwich method to determine the level of monkey rubella (Rubella) in specimens. The purified monkey rubella (Rubella) antibody was coated on the microplate to make a solid phase antibody. Rubella was added to the monoclonal antibody-coated microwells in turn, and then combined with HRP-labeled rubella antibody to form an antibody-antigen Enzyme-labeled antibody complexes were washed thoroughly and added with substrate TMB for color development. TMB is converted into blue under the catalysis of HRP enzyme, and into the final yellow under the action of acid. The color depth is positively correlated with rubella in the sample. The absorbance (OD value) was measured with a microplate reader at a wavelength of 450 nm, and the concentration of monkey rubella (Rubella) in the sample was calculated by a standard curve.
Kit composition
1 30 times concentrated washing solution 20ml × 1 bottle 7 stop solution 6ml × 1 bottle
2 Enzyme label reagent 6ml × 1 bottle 8 standard products (80IU / ml) 0.5ml × 1 bottle
3 Enzyme label coating plate 12 wells × 8 strips 9 standard dilutions 1.5ml × 1 bottle
4 Sample diluent 6ml × 1 bottle 10 instructions 1 copy
5 Developer A solution 6ml × 1 bottle 11 2 sealing film
6 Developer B liquid 6ml × 1 / bottle 12 Sealed bag 1 specimen requirement
1. Specimens are extracted as soon as possible after collection, and extraction is performed according to relevant literature, and experiments should be conducted as soon as possible after extraction. If the test cannot be performed immediately, the specimen can be stored at -20 ℃, but repeated freezing and thawing should be avoided
2. The samples containing NaN3 cannot be detected because NaN3 inhibits horseradish peroxidase (HRP) activity.
Steps
1. Dilution of standard products: This kit provides one original standard product. The user can perform dilution in a small test tube according to the following chart.
40 IU / ml No. 5 standard, 150 μl primary standard, add 150 μl standard dilution
20 IU / ml No. 4 standard 150 μl No. 5 standard added 150 μl standard dilution
10 IU / ml No. 3 standard 150μl No. 4 standard added 150μl standard dilution
5 IU / ml No. 2 standard 150μl No. 3 standard added 150μl standard diluent
2.5 IU / ml No. 1 standard 150μl No. 2 standard added 150μl standard dilution


2. Sample addition: set up blank wells (the blank control wells do not add samples and enzyme reagents, the rest of the operation is the same), standard wells, sample wells to be tested. Accurately add 50μl of the standard on the enzyme-coated plate, add 40μl of sample diluent to the sample well, and then add 10μl of the sample to be tested (the final dilution of the sample is 5 times) Add the sample and add the sample to the bottom of the well of the microplate, try not to touch the wall of the well, shake gently to mix.
3. Incubation: Seal the plate with a sealing plate and incubate at 37 ° C for 30 minutes.
4. Mixing solution: dilute 20 times concentrated washing liquid with distilled water 20 times and reserve
5. Washing: Carefully peel off the sealing film, discard the liquid, spin dry, fill each well with the washing liquid, let it stand for 30 seconds and then discard, repeat this 5 times, pat dry.
6. Add enzyme: add 50μl of enzyme label reagent to each well, except blank well.
7. Incubation: The operation is the same as 3.
8. Washing: The operation is the same as 5.
9. Color development: add 50μl of developer A to each well, then add 50μl of developer B, mix gently, and develop at 37 ° C in the dark for 15 minutes.
10. Termination: Add 50μl of stop solution to each well to stop the reaction (at this time the blue will turn to yellow).
11. Determination: Measure the absorbance (OD value) of each well in sequence with the blank air conditioner at zero and 450 nm wavelength. The measurement should be carried out within 15 minutes after adding the stop solution.
Summary of operating procedures:
1. Prepare reagents, samples and standards
2. Add the prepared samples and standards, and react at 37 ℃ for 30 minutes
3. Wash the plate 5 times, add enzyme-labeled reagent, and react at 37 ℃ for 30 minutes
4. Wash the plate 5 times, add color developing solutions A and B, and react at 37 ℃ for 10 minutes
5. Add stop solution
OD value within 6.15 minutes
7. The calculation takes the concentration of the standard as the abscissa and the OD value as the ordinate. Draw a standard curve on the graph paper. According to the OD value of the sample, find the corresponding concentration from the standard curve; multiply by the dilution factor; or use Calculate the linear regression equation of the standard curve with the concentration and OD value of the standard. Substitute the OD value of the sample into the equation to calculate the sample concentration and multiply it by the dilution factor to obtain the actual concentration of the sample.
Precautions
1. The kit should be equilibrated at room temperature for 15-30 minutes before being taken out of the refrigerated environment. If the enzyme label coated plate is unopened, the strip should be stored in a sealed bag.
2. Crystals may be precipitated in the concentrated washing liquid, which can be heated and dissolved in a water bath during dilution, and the results will not be affected during washing.
3. The sampler should be used at each step of sample addition, and the accuracy should be regularly checked to avoid test errors. It is best to control the sampling time within 5 minutes. If there are many specimens, it is recommended to use a volley gun to add samples.
4. Please make a standard curve at the same time of each measurement, it is best to make a double hole. If the content of the test substance in the specimen is too high (the OD value of the sample is greater than the OD value of the first well of the standard well), please first dilute it with a certain multiple (n times) of the sample diluent and then determine it. When calculating, please multiply the total dilution Multiple (× n × 5).
5. The sealing film is limited to one-time use to avoid cross-contamination.
6. Please keep the substrate away from light.
7. Strictly follow the instructions, and the test results must be determined by the microplate reader.
8. All samples, washing liquids and various wastes should be treated as infectious agents.
9. The components of different batches of this reagent shall not be mixed.
10. If there is any difference with the English manual, the English manual shall prevail.
Storage conditions and validity period
1. Kit storage :; 2-8 ℃.
2. Validity: 6 months



we have many kinds aprons for you reference,disposable and reusable apron.Our apron can be waterproof and oil-proof. 

Normally using for housework/meat industry and so on

We also offer OEM service so other pattern/size can be available!

reusable aprondisposable apron


. If your design of product is ready, send us, we can offer you the best price and the best service. And once you choose us and place us order, we do finish your order on time and deliver the goods to you.

Trust us, we have the best team and enough experience in the textile exporting business. We can offer you the best everything. If you are interesting on our any products, please feel free to contact us. 


Plastic Apron

Plastic Apron,PVC Apron,PE Disposable Apron,Reusable Plastic Apron

Zhaoxian Tengxing Plastic Fashion Co., Ltd , https://www.goodraincoat.com